WLSH logo
 

Environmental Health Division home page

Detection Capabilities for ICP-MS (EPA Method 1638)

Metal
MDL ppb
LOQ ppb
ML
ppb
Recommended
Analytical m/z
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Chromium
Cobalt
Copper
Lead
Manganese
Molybdenum
Nickel
Selenium
Silver
Thallium
Thorium
Uranium
Vanadium
Zinc
0.1
0.01
0.1
0.01
0.01
0.01
0.02
0.005
0.01
0.005
0.01
0.008
0.09
0.3
0.009
0.004
0.01
0.004
0.01
0.04
0.3
0.03
0.4
0.03
0.04
0.03
0.06
0.017
0.04
0.015
0.03
0.025
0.30
1.1
0.028
0.012
0.03
0.014
0.03
0.1
0.5
0.05
0.5
0.05
0.05
0.05
0.05
0.02
0.05
0.02
0.05
0.02
0.2
1.0
0.02
0.01
0.05
0.01
0.05
0.1
27
121
75
138
9
114
52
59
63
206, 207, 208
55
98
60
82
107
205
232
238
51
66

 

Notes

1. Method Detection Limit as determined by 40 CFR Part 136, Appendix B.

2. Minimum Level (ML) calculated by multiplying laboratory-determined MDL by 3.18 and rounding result to nearest multiple of 1, 2, 5, 10, etc. in accordance with procedures used by EAD and described in the EPA Draft National Guidance for the Permitting, Monitoring, and Enforcement of Water Quality-Based Effluent Limitations Set Below Analytical Detection/Quantitation Levels, March 22, 1994.

 

 
Wisconsin State Laboratory of Hygiene
 
Clinical Laboratories
465 Henry Mall
Madison WI 53706
Ph: 608-262-1293
1-888-494-4324
Fax: 1-608-262-3257
Environmental Laboratories
2601 Agriculture Drive
Madison WI 53718
Ph: 608-224-6202
1-800-442-4618
Fax: 1-608-224-6213
Walton Commons
2810 Walton Commons West
Madison, WI 53718
Help Desk: 608-221-6262
Fax: 1-608-221-6297

Copyright © 2007 Board of Regents, University of Wisconsin System
Locations | Contact | Legal Notices | Acceptable Use | Privacy Policy

Feedback, questions or accessibility issues: webmaster@mail.slh.wisc.edu
Last Update: 6:9:15 PM CST, Thursday February 1 2007